Petrac D
Electrical Therapy of Atrial Fibrillation
Journal of Clinical and Basic Cardiology 2001; 4 (2): 123-129

PDF    Summary    Overview   

Fig. 1: Vorhofflimmern - Stabiliserung Fig. 2: Jewel AF 7250 This Image - Fig. 3: Vorhofflimmern - Antitachykardes Pacing
Figure/Graphic 3: Vorhofflimmern - Antitachykardes Pacing
Termination of atrial tachycardia by antitachycardia pacing. In the upper panel, the atrial tachycardia with a mean cycle length of 440 ms is starting. After detection of tachycardia, the Jewel AF ICD starts with an atrial burst pacing (bottom panel) and terminates tachycardia. AS = atrial sensing, TS = tachycardia sensing, VP = ventricular pacing, TD = tachycardia detection, AP = atrial pacing
 
Vorhofflimmern - Antitachykardes Pacing
Previous Image Next Image  


Figure/Graphic 3: Vorhofflimmern - Antitachykardes Pacing
Termination of atrial tachycardia by antitachycardia pacing. In the upper panel, the atrial tachycardia with a mean cycle length of 440 ms is starting. After detection of tachycardia, the Jewel AF ICD starts with an atrial burst pacing (bottom panel) and terminates tachycardia. AS = atrial sensing, TS = tachycardia sensing, VP = ventricular pacing, TD = tachycardia detection, AP = atrial pacing
 